1 November 1982 New Phantoms For Measuring Low Contrast Resolution And Comparison Of Several CT Scanners When Using Them
E. Takenaka, T. A. Iinuma, M. Inoue
Author Affiliations +
Abstract
Measurements of low contrast resolution are most important. Xyron material .( stylene grafted poly-phenyl ether made by Asahi Medical Co.) can have an arbitrary CT number near to CT number of water. Phantoms with xyron-bar-in-water were made. But these phantoms showed different CT number between several CT scanners. They had energy dependency of CT number and were not suitable to measure low con-trast resolution. Then phantoms with xyron-bar-in-xyron were made. These phantomsshowed better differences between CT scanners than the former phantoms. These phantoms are most suitable to measure performance of low contrast resolution. CT number of xyron material decreases as X-ray tube voltage decreases. But differences of xyron bar contrast do not change so much.
© (1982) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
E. Takenaka, T. A. Iinuma, and M. Inoue "New Phantoms For Measuring Low Contrast Resolution And Comparison Of Several CT Scanners When Using Them", Proc. SPIE 0372, Physics and Engineering in Medical Imaging, (1 November 1982); https://doi.org/10.1117/12.934516