3 October 2016 Design and characterization of a low profile NaI(Tl) gamma camera for dedicated molecular breast tomosynthesis
Andrew M. Polemi, Justin Niestroy, Alexander Stolin, Gangadhar Jaliparthi, Randy Wojcik, Stan Majewski, Mark B. Williams
Author Affiliations +
Abstract
A new low profile gamma camera is being developed for use in a dual modality (x-ray transmission and gamma-ray emission) tomosynthesis system. Compared to the system’s current gamma camera, the new camera has a larger field of view (~20x25 cm) to better match the system’s x-ray detector (~23x29 cm), and is thinner (7.3 cm instead of 10.3 cm) permitting easier camera positioning near the top surface of the breast. It contains a pixelated NaI(Tl) array with a crystal pitch of 2.2 mm, which is optically coupled to a 4x5 array of Hamamatsu H8500C position sensitive photomultiplier tubes (PSPMTs). The manufacturer-provided connector board of each PSPMT was replaced with a custom designed board that a) reduces the 64 channel readout of the 8x8 electrode anode of the H8500C to 16 channels (8X and 8Y), b) performs gain non-uniformity correction, and c) reduces the height of the PSPMT-base assembly, 37.7 mm to 27.87 mm. The X and Y outputs of each module are connected in a lattice framework, and at two edges of this lattice, the X and Y outputs (32Y by 40X) are coupled to an amplifier/output board whose signals are fed via shielded ribbon cables to external ADCs. The camera uses parallel hole collimation. We describe the measured camera imaging performance, including intrinsic and extrinsic spatial resolution, detection sensitivity, uniformity of response, energy resolution for 140 keV gamma rays, and geometric linearity.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Andrew M. Polemi, Justin Niestroy, Alexander Stolin, Gangadhar Jaliparthi, Randy Wojcik, Stan Majewski, and Mark B. Williams "Design and characterization of a low profile NaI(Tl) gamma camera for dedicated molecular breast tomosynthesis", Proc. SPIE 9969, Radiation Detectors: Systems and Applications XVII, 99690O (3 October 2016); https://doi.org/10.1117/12.2246561
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Cameras

Crystals

Spatial resolution

Imaging systems

Collimators

Breast

Gamma radiation

Back to Top