The DPS mechanism will be exclusively utilized during the alignment and test phase. Upon completing the test phase, it will be mechanically locked at the best pre-determined focus so that it cannot be moved during the observation period.
The DPS has been conceptualized as a fixed and reproducible interface to the Main Bench Structure in the MICADO cryostat and as an adjustable unit containing the Detector Array mounted on the DPS frame installed on a linear guide on the base plate. A cryogenic linear actuator further acts as the linear guide during the alignment phase to bring the focal plane array into focus.
One of the key features of the upgraded spectrograph is the new calibration system, which uses a laser frequency comb as reference light source. Another aspect is the possibility to perform simultaneous wavelength calibration, while recording science data. For this purpose a new 4-fiber slit has been developed, which opens up the possibility to feed light from different sources at the same time through the entrance slit of the spectrograph. We present a detailed characterization of this new device, based on the results of extensive lab tests performed at the Munich University Observatory.